An investigation of transmission efficiency of transcutaneous energy transmission system

R. Tashiro, N. Kabei, K. Tsuchiya

Research output: Contribution to journalArticle

Abstract

To supply an electric power for driving a total artificial heart (TAH) from outside of the body, transcutaneous energy transmission system is profitable, but energy transmission losses occur. Especially, it is considered to be easily affected by forms and/or disposition of two coils. In this study, we have made a DC-AC inverter, rectifying and smoothing circuit and coreless coils, and experimentally investigated the influence of total energy transmission efficiency caused by vertical and/or horizontal disposition of two coils. In case of two same coils, maximum efficiency is obtained, but transmission stability for a horizontal shift is inferior. In case of large and small coils, for constant resistance of a load, energy supply, energy consumption and total transmission efficiency are all stable when an air core of a large cell contains an air core of a small one.

Original languageEnglish
Pages (from-to)293-297
Number of pages5
JournalJapanese Journal of Artificial Organs
Volume26
Issue number2
Publication statusPublished - 1997
Externally publishedYes

Keywords

  • total artificial heart
  • transcutaneous energy transmission

ASJC Scopus subject areas

  • Biophysics

Cite this

An investigation of transmission efficiency of transcutaneous energy transmission system. / Tashiro, R.; Kabei, N.; Tsuchiya, K.

In: Japanese Journal of Artificial Organs, Vol. 26, No. 2, 1997, p. 293-297.

Research output: Contribution to journalArticle

Tashiro, R. ; Kabei, N. ; Tsuchiya, K. / An investigation of transmission efficiency of transcutaneous energy transmission system. In: Japanese Journal of Artificial Organs. 1997 ; Vol. 26, No. 2. pp. 293-297.
@article{25d6eee81bff4c5d9fbd12d5de147b87,
title = "An investigation of transmission efficiency of transcutaneous energy transmission system",
abstract = "To supply an electric power for driving a total artificial heart (TAH) from outside of the body, transcutaneous energy transmission system is profitable, but energy transmission losses occur. Especially, it is considered to be easily affected by forms and/or disposition of two coils. In this study, we have made a DC-AC inverter, rectifying and smoothing circuit and coreless coils, and experimentally investigated the influence of total energy transmission efficiency caused by vertical and/or horizontal disposition of two coils. In case of two same coils, maximum efficiency is obtained, but transmission stability for a horizontal shift is inferior. In case of large and small coils, for constant resistance of a load, energy supply, energy consumption and total transmission efficiency are all stable when an air core of a large cell contains an air core of a small one.",
keywords = "total artificial heart, transcutaneous energy transmission",
author = "R. Tashiro and N. Kabei and K. Tsuchiya",
year = "1997",
language = "English",
volume = "26",
pages = "293--297",
journal = "Japanese Journal of Artificial Organs",
issn = "0300-0818",
publisher = "Japanese Society for Artificial Organs and Tissues",
number = "2",

}

TY - JOUR

T1 - An investigation of transmission efficiency of transcutaneous energy transmission system

AU - Tashiro, R.

AU - Kabei, N.

AU - Tsuchiya, K.

PY - 1997

Y1 - 1997

N2 - To supply an electric power for driving a total artificial heart (TAH) from outside of the body, transcutaneous energy transmission system is profitable, but energy transmission losses occur. Especially, it is considered to be easily affected by forms and/or disposition of two coils. In this study, we have made a DC-AC inverter, rectifying and smoothing circuit and coreless coils, and experimentally investigated the influence of total energy transmission efficiency caused by vertical and/or horizontal disposition of two coils. In case of two same coils, maximum efficiency is obtained, but transmission stability for a horizontal shift is inferior. In case of large and small coils, for constant resistance of a load, energy supply, energy consumption and total transmission efficiency are all stable when an air core of a large cell contains an air core of a small one.

AB - To supply an electric power for driving a total artificial heart (TAH) from outside of the body, transcutaneous energy transmission system is profitable, but energy transmission losses occur. Especially, it is considered to be easily affected by forms and/or disposition of two coils. In this study, we have made a DC-AC inverter, rectifying and smoothing circuit and coreless coils, and experimentally investigated the influence of total energy transmission efficiency caused by vertical and/or horizontal disposition of two coils. In case of two same coils, maximum efficiency is obtained, but transmission stability for a horizontal shift is inferior. In case of large and small coils, for constant resistance of a load, energy supply, energy consumption and total transmission efficiency are all stable when an air core of a large cell contains an air core of a small one.

KW - total artificial heart

KW - transcutaneous energy transmission

UR - http://www.scopus.com/inward/record.url?scp=0030951670&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030951670&partnerID=8YFLogxK

M3 - Article

VL - 26

SP - 293

EP - 297

JO - Japanese Journal of Artificial Organs

JF - Japanese Journal of Artificial Organs

SN - 0300-0818

IS - 2