Abstract

Even though blood pumps have come into clinical usage, thrombo-embolic complications still pose a major problem, and they have not yet been clarified and quantified. However, it is known that the basis of thrombus formation is platelet adhesion, which is thought to be closely associated with the shear rate. Therefore, our current interest focuses on the effect of shear conditions on platelet adhesion. We have designed and carried out an experimental setup allowing fluorescent microscopy of whole blood within a rotational viscometer under controllable shear conditions. A small area of the bottom plate was coated with type I collagen, which provided a model of the injured vessel as a target for platelet adhesion. Using this setup, the time course of platelet adhesion under several different shear rates, ranging from 127 to 723 s -1, was studied. Platelet adhesion increased along with shear rates up to 283 s -1, followed by a gradual decrease when the shear rate exceeded 346 s -1. The adhesion amounts were statistically significant between 283 and 173 s -1 (p = 0. 02), 173 and 127 s -1 (p = 0. 035), and 283 and 503 s -1 (p = 0. 03), respectively. This result suggests that there is an optimal shear condition around 300 s -1 for platelet adhesion to type I collagen.

Original languageEnglish
Pages (from-to)64-70
Number of pages7
JournalJournal of Biorheology
Volume25
Issue number1-2
DOIs
Publication statusPublished - 2011 Dec

Fingerprint

Rotational flow
Platelets
Adhesion
Shear deformation
Collagen Type I
Collagen
Blood
Viscometers
Microscopic examination
Pumps

Keywords

  • Collagen-coated surface
  • Initial thrombus formation
  • Microscopic observation
  • Shear rate

ASJC Scopus subject areas

  • Mechanics of Materials
  • Materials Science(all)
  • Mechanical Engineering

Cite this

Watanabe, N., Affeld, K., Schaller, J., Schmitmeier, S., Reininger, A. J., Goubergrits, L., & Kertzscher, U. (2011). Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber. Journal of Biorheology, 25(1-2), 64-70. https://doi.org/10.1007/s12573-011-0039-y

Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber. / Watanabe, Nobuo; Affeld, Klaus; Schaller, Jens; Schmitmeier, Stephanie; Reininger, Armin J.; Goubergrits, Leonid; Kertzscher, Ulrich.

In: Journal of Biorheology, Vol. 25, No. 1-2, 12.2011, p. 64-70.

Research output: Contribution to journalArticle

Watanabe, N, Affeld, K, Schaller, J, Schmitmeier, S, Reininger, AJ, Goubergrits, L & Kertzscher, U 2011, 'Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber', Journal of Biorheology, vol. 25, no. 1-2, pp. 64-70. https://doi.org/10.1007/s12573-011-0039-y
Watanabe, Nobuo ; Affeld, Klaus ; Schaller, Jens ; Schmitmeier, Stephanie ; Reininger, Armin J. ; Goubergrits, Leonid ; Kertzscher, Ulrich. / Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber. In: Journal of Biorheology. 2011 ; Vol. 25, No. 1-2. pp. 64-70.
@article{f4458a200a4b471dafa1996942cbce1a,
title = "Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber",
abstract = "Even though blood pumps have come into clinical usage, thrombo-embolic complications still pose a major problem, and they have not yet been clarified and quantified. However, it is known that the basis of thrombus formation is platelet adhesion, which is thought to be closely associated with the shear rate. Therefore, our current interest focuses on the effect of shear conditions on platelet adhesion. We have designed and carried out an experimental setup allowing fluorescent microscopy of whole blood within a rotational viscometer under controllable shear conditions. A small area of the bottom plate was coated with type I collagen, which provided a model of the injured vessel as a target for platelet adhesion. Using this setup, the time course of platelet adhesion under several different shear rates, ranging from 127 to 723 s -1, was studied. Platelet adhesion increased along with shear rates up to 283 s -1, followed by a gradual decrease when the shear rate exceeded 346 s -1. The adhesion amounts were statistically significant between 283 and 173 s -1 (p = 0. 02), 173 and 127 s -1 (p = 0. 035), and 283 and 503 s -1 (p = 0. 03), respectively. This result suggests that there is an optimal shear condition around 300 s -1 for platelet adhesion to type I collagen.",
keywords = "Collagen-coated surface, Initial thrombus formation, Microscopic observation, Shear rate",
author = "Nobuo Watanabe and Klaus Affeld and Jens Schaller and Stephanie Schmitmeier and Reininger, {Armin J.} and Leonid Goubergrits and Ulrich Kertzscher",
year = "2011",
month = "12",
doi = "10.1007/s12573-011-0039-y",
language = "English",
volume = "25",
pages = "64--70",
journal = "Journal of Biorheology",
issn = "1867-0466",
publisher = "Springer Japan",
number = "1-2",

}

TY - JOUR

T1 - Investigation of human platelet adhesion under low shear conditions in a rotational flow chamber

AU - Watanabe, Nobuo

AU - Affeld, Klaus

AU - Schaller, Jens

AU - Schmitmeier, Stephanie

AU - Reininger, Armin J.

AU - Goubergrits, Leonid

AU - Kertzscher, Ulrich

PY - 2011/12

Y1 - 2011/12

N2 - Even though blood pumps have come into clinical usage, thrombo-embolic complications still pose a major problem, and they have not yet been clarified and quantified. However, it is known that the basis of thrombus formation is platelet adhesion, which is thought to be closely associated with the shear rate. Therefore, our current interest focuses on the effect of shear conditions on platelet adhesion. We have designed and carried out an experimental setup allowing fluorescent microscopy of whole blood within a rotational viscometer under controllable shear conditions. A small area of the bottom plate was coated with type I collagen, which provided a model of the injured vessel as a target for platelet adhesion. Using this setup, the time course of platelet adhesion under several different shear rates, ranging from 127 to 723 s -1, was studied. Platelet adhesion increased along with shear rates up to 283 s -1, followed by a gradual decrease when the shear rate exceeded 346 s -1. The adhesion amounts were statistically significant between 283 and 173 s -1 (p = 0. 02), 173 and 127 s -1 (p = 0. 035), and 283 and 503 s -1 (p = 0. 03), respectively. This result suggests that there is an optimal shear condition around 300 s -1 for platelet adhesion to type I collagen.

AB - Even though blood pumps have come into clinical usage, thrombo-embolic complications still pose a major problem, and they have not yet been clarified and quantified. However, it is known that the basis of thrombus formation is platelet adhesion, which is thought to be closely associated with the shear rate. Therefore, our current interest focuses on the effect of shear conditions on platelet adhesion. We have designed and carried out an experimental setup allowing fluorescent microscopy of whole blood within a rotational viscometer under controllable shear conditions. A small area of the bottom plate was coated with type I collagen, which provided a model of the injured vessel as a target for platelet adhesion. Using this setup, the time course of platelet adhesion under several different shear rates, ranging from 127 to 723 s -1, was studied. Platelet adhesion increased along with shear rates up to 283 s -1, followed by a gradual decrease when the shear rate exceeded 346 s -1. The adhesion amounts were statistically significant between 283 and 173 s -1 (p = 0. 02), 173 and 127 s -1 (p = 0. 035), and 283 and 503 s -1 (p = 0. 03), respectively. This result suggests that there is an optimal shear condition around 300 s -1 for platelet adhesion to type I collagen.

KW - Collagen-coated surface

KW - Initial thrombus formation

KW - Microscopic observation

KW - Shear rate

UR - http://www.scopus.com/inward/record.url?scp=84655167154&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84655167154&partnerID=8YFLogxK

U2 - 10.1007/s12573-011-0039-y

DO - 10.1007/s12573-011-0039-y

M3 - Article

VL - 25

SP - 64

EP - 70

JO - Journal of Biorheology

JF - Journal of Biorheology

SN - 1867-0466

IS - 1-2

ER -