Analysis of pressure distribution in transfemoral prosthetic socket for prefabrication evaluation via the finite element method

Mohd Syahmi Jamaludin, Akihiko Hanafusa, Yamamoto Shinichirou, Yukio Agarie, Hiroshi Otsuka, Kengo Ohnishi

研究成果: Article

抄録

In this study, we estimated and validated the pressure distribution profile between the residuum and two types of prosthetic sockets for transfemoral amputees by utilizing a finite element analysis. Correct shaping of the socket for an appropriate load distribution is a critical process in the design of lower-limb prosthesis sockets. The pressure distribution profile provides an understanding of the relationship between the socket design and the level of subject comfortability. Estimating the pressure profile is important, as it helps improve the prosthesis through an evaluation of the socket design before it undergoes the fabrication process. This study focused on utilizing a magnetic resonance imaging (MRI)-based three-dimensional (3D) model inside a predetermined finite element simulation. The simulation was predetermined by mimicking the actual socket-fitting environment. The results showed that the potential MRI-based 3D model simulation could be used as an estimation tool for a pressure distribution profile due to the high correlation co-efficient value (R2 > 0.8) calculated when the pressure profiles were compared to the experiment data. The simulation also showed that the pressure distribution in the proximal area was higher (~30%) than in the distal area of the prosthetic socket for every subject. The results of this study will be of tremendous interest for fabricators through the use of a finite element model as an alternative method for the prefabrication and evaluation of prosthetic sockets. In future prosthetic socket fabrications, less intervention will be required in the development of a socket, and the participation of the subject in the socket-fitting session will not be necessary. The results suggest that this study will contribute to expanding the development of an overall prefabrication evaluation system to allow healthcare providers and engineers to simulate the fit and comfort of transfemoral prosthetics.

元の言語English
記事番号98
ジャーナルBioengineering
6
発行部数4
DOI
出版物ステータスPublished - 2019 12

Fingerprint

Prosthetics
Pressure distribution
Finite element method
Magnetic resonance
Imaging techniques
Fabrication
Engineers
Experiments
Prostheses and Implants

ASJC Scopus subject areas

  • Bioengineering

これを引用

Analysis of pressure distribution in transfemoral prosthetic socket for prefabrication evaluation via the finite element method. / Jamaludin, Mohd Syahmi; Hanafusa, Akihiko; Shinichirou, Yamamoto; Agarie, Yukio; Otsuka, Hiroshi; Ohnishi, Kengo.

:: Bioengineering, 巻 6, 番号 4, 98, 12.2019.

研究成果: Article

@article{21e3c83f56544b3884d279008b99d578,
title = "Analysis of pressure distribution in transfemoral prosthetic socket for prefabrication evaluation via the finite element method",
abstract = "In this study, we estimated and validated the pressure distribution profile between the residuum and two types of prosthetic sockets for transfemoral amputees by utilizing a finite element analysis. Correct shaping of the socket for an appropriate load distribution is a critical process in the design of lower-limb prosthesis sockets. The pressure distribution profile provides an understanding of the relationship between the socket design and the level of subject comfortability. Estimating the pressure profile is important, as it helps improve the prosthesis through an evaluation of the socket design before it undergoes the fabrication process. This study focused on utilizing a magnetic resonance imaging (MRI)-based three-dimensional (3D) model inside a predetermined finite element simulation. The simulation was predetermined by mimicking the actual socket-fitting environment. The results showed that the potential MRI-based 3D model simulation could be used as an estimation tool for a pressure distribution profile due to the high correlation co-efficient value (R2 > 0.8) calculated when the pressure profiles were compared to the experiment data. The simulation also showed that the pressure distribution in the proximal area was higher (~30{\%}) than in the distal area of the prosthetic socket for every subject. The results of this study will be of tremendous interest for fabricators through the use of a finite element model as an alternative method for the prefabrication and evaluation of prosthetic sockets. In future prosthetic socket fabrications, less intervention will be required in the development of a socket, and the participation of the subject in the socket-fitting session will not be necessary. The results suggest that this study will contribute to expanding the development of an overall prefabrication evaluation system to allow healthcare providers and engineers to simulate the fit and comfort of transfemoral prosthetics.",
keywords = "3D model, Biomechanics, Finite element analysis, Finite element method, Prosthetic socket, Transfemoral residuum",
author = "Jamaludin, {Mohd Syahmi} and Akihiko Hanafusa and Yamamoto Shinichirou and Yukio Agarie and Hiroshi Otsuka and Kengo Ohnishi",
year = "2019",
month = "12",
doi = "10.3390/bioengineering6040098",
language = "English",
volume = "6",
journal = "Bioengineering",
issn = "2306-5354",
publisher = "MDPI AG",
number = "4",

}

TY - JOUR

T1 - Analysis of pressure distribution in transfemoral prosthetic socket for prefabrication evaluation via the finite element method

AU - Jamaludin, Mohd Syahmi

AU - Hanafusa, Akihiko

AU - Shinichirou, Yamamoto

AU - Agarie, Yukio

AU - Otsuka, Hiroshi

AU - Ohnishi, Kengo

PY - 2019/12

Y1 - 2019/12

N2 - In this study, we estimated and validated the pressure distribution profile between the residuum and two types of prosthetic sockets for transfemoral amputees by utilizing a finite element analysis. Correct shaping of the socket for an appropriate load distribution is a critical process in the design of lower-limb prosthesis sockets. The pressure distribution profile provides an understanding of the relationship between the socket design and the level of subject comfortability. Estimating the pressure profile is important, as it helps improve the prosthesis through an evaluation of the socket design before it undergoes the fabrication process. This study focused on utilizing a magnetic resonance imaging (MRI)-based three-dimensional (3D) model inside a predetermined finite element simulation. The simulation was predetermined by mimicking the actual socket-fitting environment. The results showed that the potential MRI-based 3D model simulation could be used as an estimation tool for a pressure distribution profile due to the high correlation co-efficient value (R2 > 0.8) calculated when the pressure profiles were compared to the experiment data. The simulation also showed that the pressure distribution in the proximal area was higher (~30%) than in the distal area of the prosthetic socket for every subject. The results of this study will be of tremendous interest for fabricators through the use of a finite element model as an alternative method for the prefabrication and evaluation of prosthetic sockets. In future prosthetic socket fabrications, less intervention will be required in the development of a socket, and the participation of the subject in the socket-fitting session will not be necessary. The results suggest that this study will contribute to expanding the development of an overall prefabrication evaluation system to allow healthcare providers and engineers to simulate the fit and comfort of transfemoral prosthetics.

AB - In this study, we estimated and validated the pressure distribution profile between the residuum and two types of prosthetic sockets for transfemoral amputees by utilizing a finite element analysis. Correct shaping of the socket for an appropriate load distribution is a critical process in the design of lower-limb prosthesis sockets. The pressure distribution profile provides an understanding of the relationship between the socket design and the level of subject comfortability. Estimating the pressure profile is important, as it helps improve the prosthesis through an evaluation of the socket design before it undergoes the fabrication process. This study focused on utilizing a magnetic resonance imaging (MRI)-based three-dimensional (3D) model inside a predetermined finite element simulation. The simulation was predetermined by mimicking the actual socket-fitting environment. The results showed that the potential MRI-based 3D model simulation could be used as an estimation tool for a pressure distribution profile due to the high correlation co-efficient value (R2 > 0.8) calculated when the pressure profiles were compared to the experiment data. The simulation also showed that the pressure distribution in the proximal area was higher (~30%) than in the distal area of the prosthetic socket for every subject. The results of this study will be of tremendous interest for fabricators through the use of a finite element model as an alternative method for the prefabrication and evaluation of prosthetic sockets. In future prosthetic socket fabrications, less intervention will be required in the development of a socket, and the participation of the subject in the socket-fitting session will not be necessary. The results suggest that this study will contribute to expanding the development of an overall prefabrication evaluation system to allow healthcare providers and engineers to simulate the fit and comfort of transfemoral prosthetics.

KW - 3D model

KW - Biomechanics

KW - Finite element analysis

KW - Finite element method

KW - Prosthetic socket

KW - Transfemoral residuum

UR - http://www.scopus.com/inward/record.url?scp=85074387865&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85074387865&partnerID=8YFLogxK

U2 - 10.3390/bioengineering6040098

DO - 10.3390/bioengineering6040098

M3 - Article

AN - SCOPUS:85074387865

VL - 6

JO - Bioengineering

JF - Bioengineering

SN - 2306-5354

IS - 4

M1 - 98

ER -