抄録
A unique experimental attempt aiming to obtain a uniform magnetic field space as required for NMR has been carried out with use of HTS bulk magnets. The magnetic poles were activated as 1.8 T (North) and 1.4 T (South) at 30 K by applyinga pulsed magnetic field up to 7 T, and positioned face-to-face with gaps less than 70 mm. The uniformity of the magnetic field required for detecting the NMR signals isless than 1,500 ppm at more than 0.3 T in the cross sectional plane of 2 x 2 mm2. After thepreliminary trials whichrevealed auniformity of 5,421 ppm at 0.44 T in a70 mm gap, we attached a ferromagnetic iron plate to a magnetic pole surface to change the magnetic field distribution to be concave. The best uniformity of 358 ppm at 1.11 T was obtained at 9 mm distance from the iron plate surface in a gap of 30 mm. It is stated that the concave magnetic field distribution was compensated by the counter conical-shape field, resulting in the uniform field plane.
本文言語 | English |
---|---|
ページ(範囲) | 1005-1009 |
ページ数 | 5 |
ジャーナル | Physics Procedia |
巻 | 67 |
DOI | |
出版ステータス | Published - 2015 |
外部発表 | はい |
イベント | 25th International Cryogenic Engineering Conference and International Cryogenic Materials Conference, ICEC/ICMC 2014 - Enschede, Netherlands 継続期間: 2014 7月 7 → 2014 7月 11 |
ASJC Scopus subject areas
- 物理学および天文学(全般)