C-Mode Scan and Resolution Improvement Techniques for Ultrasonic Diagnosis

Ken Ichi Ito, Masayasu Itoh, Shin Ichi Yuta, Hiromu Yokoi, Tamon Inouye

研究成果査読

2 被引用数 (Scopus)

抄録

This paper presents a new method for obtaining plane images perpendicular to the axis of the transducer and discusses some resolution improvement techniques utilizing a computer. The method, called C-mode scan, is carried out by successive application of a conventional B-mode scan, and supplies more information such as the outline, size, and extent of a lesion such as a breast tumor. To improve the resolution of a transducer, a deconvolution technique together with non-linear filtering are employed from a spectral point of view. Several clinical images produced by C-mode and the computer improvement of resolution are demonstrated.

本文言語English
ページ(範囲)11-17
ページ数7
ジャーナルIEEE Transactions on Biomedical Engineering
BME-26
1
DOI
出版ステータスPublished - 1979 1月

ASJC Scopus subject areas

  • 生体医工学

フィンガープリント

「C-Mode Scan and Resolution Improvement Techniques for Ultrasonic Diagnosis」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。