Determination of pore size distribution of dialysis membranes from solute permeability data

Y. Yoshimi, W. Okada, K. Sakai

研究成果: Article査読

抄録

Pore size distribution was determined from solute permeability data for urea, glucose and sucrose as a function of Strokes radius. Regenerated cellulose membranes have more limited pore size distribution than polymethylmethacrylate membrane. The solute permeability of glucose oligomers (monomer to octomer) was calculated from clearance data obtained by dialysis experiments.

本文言語English
ページ(範囲)675-678
ページ数4
ジャーナルJapanese Journal of Artificial Organs
19
2
出版ステータスPublished - 1990 1 1