Microfluidic polymer chip with an embedded ion-selective electrode detector for nitrate-ion assay in environmental samples

Takashi Masadome, Kazuki Nakamura, Daisuke Iijima, Osamu Horiuchi, Benjaporn Tossanaitada, Shin ichi Wakida, Toshihiko Imato

研究成果: Article査読

11 被引用数 (Scopus)

抄録

A nitrate ion-selective electrode (NO 3 - - ISE) has been developed based on tetradodecylammonium bromide as an anion exchanger and 2-nitrophenyl octyl ether as a plasticizer. The NO 3 - - ISE shows an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10 -6 and 1.0 × 10 -1 M, with an anionic slope of -57.7 ± 0.7 mV/decade. The selectivity coefficients of the NO 3 - - ISE for nitrate ion against chloride and sulfate were obtained. A microfluidic polymer chip was fabricated using polystyrene plates and stainless-steel wires as a template for the channel. The microfluidic polymer chip is composed of a mixing chip and a NO 3 - - ISE detector chip. The microfluidic polymer chip, integrated with a NO 3 - - ISE detector consisting of the NO 3 - - ISE and a Na + -ISE as a reference electrode, showed an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10 -5 and 1.0 × 10 -1 M, with an anionic slope of -54.3 ± 1.3 mV/decade. The microfluidic polymer chip was then applied to the determination of nitrate ion in environmental water samples, such as a tap water, a well-water sample and water for agricultural use. 2010

本文言語English
ページ(範囲)417-423
ページ数7
ジャーナルAnalytical Sciences
26
4
DOI
出版ステータスPublished - 2010 4

ASJC Scopus subject areas

  • 分析化学

フィンガープリント

「Microfluidic polymer chip with an embedded ion-selective electrode detector for nitrate-ion assay in environmental samples」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル