New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 2: correlation study

K. Yamakoshi, H. Shimazu, Masahiro Shibata, A. Kamiya

研究成果: Article

87 引用 (Scopus)

抄録

With a new oscillometric method equipped with a transmittance infrared photoelectric plethysmograph, indirect systolic, and mean arterial pressures were measured in 12 normotensive and hypertensive subjects with systolic and mean arterial pressures ranging from 55 to 163 mmHg and from 95 to 200 mmHg, respectively. The pressure values obtained by this method were compared with direct measurements of the brachial intra-arterial pressure recorded simultaneously. A fairly good correlation between the pairs of simultaneous data from these two methods were obtained.

元の言語English
ページ(範囲)314-318
ページ数5
ジャーナルMedical & Biological Engineering & Computing
20
発行部数3
DOI
出版物ステータスPublished - 1982 5
外部発表Yes

Fingerprint

Infrared radiation

ASJC Scopus subject areas

  • Health Information Management
  • Health Informatics
  • Biomedical Engineering
  • Computer Science Applications
  • Computational Theory and Mathematics

これを引用

@article{0dd4e813694c455f90598d0072d42804,
title = "New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 2: correlation study",
abstract = "With a new oscillometric method equipped with a transmittance infrared photoelectric plethysmograph, indirect systolic, and mean arterial pressures were measured in 12 normotensive and hypertensive subjects with systolic and mean arterial pressures ranging from 55 to 163 mmHg and from 95 to 200 mmHg, respectively. The pressure values obtained by this method were compared with direct measurements of the brachial intra-arterial pressure recorded simultaneously. A fairly good correlation between the pairs of simultaneous data from these two methods were obtained.",
keywords = "Finger arterial pressure, Indirect systolic and mean pressure, Oscillometric method, Photoelectric plethysmography",
author = "K. Yamakoshi and H. Shimazu and Masahiro Shibata and A. Kamiya",
year = "1982",
month = "5",
doi = "10.1007/BF02442798",
language = "English",
volume = "20",
pages = "314--318",
journal = "Medical and Biological Engineering and Computing",
issn = "0140-0118",
publisher = "Springer Verlag",
number = "3",

}

TY - JOUR

T1 - New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 2

T2 - correlation study

AU - Yamakoshi, K.

AU - Shimazu, H.

AU - Shibata, Masahiro

AU - Kamiya, A.

PY - 1982/5

Y1 - 1982/5

N2 - With a new oscillometric method equipped with a transmittance infrared photoelectric plethysmograph, indirect systolic, and mean arterial pressures were measured in 12 normotensive and hypertensive subjects with systolic and mean arterial pressures ranging from 55 to 163 mmHg and from 95 to 200 mmHg, respectively. The pressure values obtained by this method were compared with direct measurements of the brachial intra-arterial pressure recorded simultaneously. A fairly good correlation between the pairs of simultaneous data from these two methods were obtained.

AB - With a new oscillometric method equipped with a transmittance infrared photoelectric plethysmograph, indirect systolic, and mean arterial pressures were measured in 12 normotensive and hypertensive subjects with systolic and mean arterial pressures ranging from 55 to 163 mmHg and from 95 to 200 mmHg, respectively. The pressure values obtained by this method were compared with direct measurements of the brachial intra-arterial pressure recorded simultaneously. A fairly good correlation between the pairs of simultaneous data from these two methods were obtained.

KW - Finger arterial pressure

KW - Indirect systolic and mean pressure

KW - Oscillometric method

KW - Photoelectric plethysmography

UR - http://www.scopus.com/inward/record.url?scp=0020137931&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020137931&partnerID=8YFLogxK

U2 - 10.1007/BF02442798

DO - 10.1007/BF02442798

M3 - Article

C2 - 7109727

AN - SCOPUS:0020137931

VL - 20

SP - 314

EP - 318

JO - Medical and Biological Engineering and Computing

JF - Medical and Biological Engineering and Computing

SN - 0140-0118

IS - 3

ER -