New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 2: correlation study

K. Yamakoshi, H. Shimazu, M. Shibata, A. Kamiya

研究成果: Article

88 引用 (Scopus)

抜粋

With a new oscillometric method equipped with a transmittance infrared photoelectric plethysmograph, indirect systolic, and mean arterial pressures were measured in 12 normotensive and hypertensive subjects with systolic and mean arterial pressures ranging from 55 to 163 mmHg and from 95 to 200 mmHg, respectively. The pressure values obtained by this method were compared with direct measurements of the brachial intra-arterial pressure recorded simultaneously. A fairly good correlation between the pairs of simultaneous data from these two methods were obtained.

元の言語English
ページ(範囲)314-318
ページ数5
ジャーナルMedical & Biological Engineering & Computing
20
発行部数3
DOI
出版物ステータスPublished - 1982 5 1

ASJC Scopus subject areas

  • Biomedical Engineering
  • Computer Science Applications

フィンガープリント New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 2: correlation study' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用