抄録
A sequential injection analysis (SIA) system for the determination of NO3- ions using a microfluidic polymer chip with an embedded NO3- ion-selective electrode (NO3 --ISE) as a detector was developed. Under the optimal flow conditions of the SIA system, the NO3--ISE showed a good linear relationship between peak heights and logarithmic concentrations of NO 3- ions with a Nernstian slope of 57.1 mV per decade in a concentration range from 1.0 × 10-5 to 1.0 × 10 -1 mol dm-3. Sample throughput of the present system for the determination of NO3- ions was ca. 12 samples per h. The proposed SIA system was applied to the determination of the level of NO 3- ions in environmental water samples, such as tap water, well water and river water.
本文言語 | English |
---|---|
ページ(範囲) | 4384-4388 |
ページ数 | 5 |
ジャーナル | Analytical Methods |
巻 | 4 |
号 | 12 |
DOI | |
出版ステータス | Published - 2012 12月 |