Use of marker ion and cationic surfactant plastic membrane electrode for potentiometric titration of cationic polyelectrolytes

Takashi Masadome, Toshihiko Imato

研究成果: Article査読

9 被引用数 (Scopus)

抄録

A plasticized poly (vinyl chloride) (PVC) membrane electrode sensitive to stearyltrimethylammonium (STA) ion is applied to the determination of cationic polyelectrolytes such as poly (diallyldimethylammonium chloride) (Cat-floc) by potentiometric titration, using a potassium poly (vinyl sulfate) (PVSK) solution as a titrant. The end-point of the titration is detected as the potential change of the plasticized PVC membrane electrode caused by decrease in the concentration of STA ion added to the sample solution as a marker ion due to the ion association reaction between the STA ion and PVSK. The effects of the concentration of STA ion, coexisting electrolytes in the sample solution and pH of the sample on the degree of the potential change at the end-point were examined. A linear relationship between the concentration of cationic polyelectrolyte and the end-point volume of the titrant exists in the concentration range from 2 × 10-5 to 4 × 10-4 N for Cat-floc, glycol chitosan, and methylglycol chitosan.

本文言語English
ページ(範囲)663-668
ページ数6
ジャーナルTalanta
60
4
DOI
出版ステータスPublished - 2003 7 4
外部発表はい

ASJC Scopus subject areas

  • 分析化学

フィンガープリント

「Use of marker ion and cationic surfactant plastic membrane electrode for potentiometric titration of cationic polyelectrolytes」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル